PlainRecalls
FDA Devices Moderate Class II Ongoing

Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones, Part Number: 1067716

Reported: May 1, 2024 Initiated: March 22, 2024 #Z-1616-2024

Product Description

Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones, Part Number: 1067716

Reason for Recall

Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

Details

Recalling Firm
EBI, LLC
Units Affected
N/A
Distribution
US Nationwide distribution.
Location
Parsippany, NJ

Frequently Asked Questions

What product was recalled?
Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones, Part Number: 1067716. Recalled by EBI, LLC. Units affected: N/A.
Why was this product recalled?
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
Which agency issued this recall?
This recall was issued by the FDA Devices on May 1, 2024. Severity: Moderate. Recall number: Z-1616-2024.