Fresenius 2008T Hemodialysis Machine with Revision K Actuator Test Boards and also sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 190234 Revision K
Reported: May 21, 2014 Initiated: April 25, 2014 #Z-1618-2014
Product Description
Fresenius 2008T Hemodialysis Machine with Revision K Actuator Test Boards and also sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 190234 Revision K
Reason for Recall
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
Details
- Recalling Firm
- Fresenius Medical Care Holdings, Inc.
- Units Affected
- 79 units
- Distribution
- USA (nationwide) and the country of Canada.*
- Location
- Waltham, MA
Frequently Asked Questions
What product was recalled? ▼
Fresenius 2008T Hemodialysis Machine with Revision K Actuator Test Boards and also sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 190234 Revision K. Recalled by Fresenius Medical Care Holdings, Inc.. Units affected: 79 units.
Why was this product recalled? ▼
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 21, 2014. Severity: Moderate. Recall number: Z-1618-2014.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11