PlainRecalls
FDA Devices Moderate Class II Ongoing

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Reported: May 1, 2024 Initiated: February 28, 2024 #Z-1620-2024

Product Description

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Reason for Recall

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Details

Units Affected
444 Screw Heads
Distribution
US: CA, FL, CO, TX
Location
Carlsbad, CA

Frequently Asked Questions

What product was recalled?
Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.. Recalled by SEASPINE ORTHOPEDICS CORPORATION. Units affected: 444 Screw Heads.
Why was this product recalled?
Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 1, 2024. Severity: Moderate. Recall number: Z-1620-2024.