Severity
Moderate
FDA Devices recall · Reported May 28, 2014
The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result…
Philips Medical Systems, Inc. recalled Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The mon… - a moderate-severity action.
Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The mon… was recalled by Philips Medical Systems, Inc. in May 28, 2014. Reason: The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date …. Check the official notice for the remedy. Verify recall #Z-1628-2014 with the FDA Devices before acting.
The recall
Philips Medical Systems, Inc. issued this moderate-severity FDA Devices recall-The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1628-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1628-2014) was formally reported on May 28, 2014, with the manufacturer initiating the action on April 4, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems, Inc. is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 188,520 devices, a scale large enough to require multi-state distribution tracking.
The documented reason for this recall is: The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can c… Distribution data in the federal record shows the product reached: Worldwide Distribution - Including US Nationwide, Canada, Czech Republic, Denmark, Netherlands, Estonia, Finland, France, Germany, Greece, Hungary, Italy, Japan, Kazakhstan, Norway, Poland, Romania, Russia, China, Slova…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
188,520 devices
Related Recalls
6
3 from same agency
Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.
The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1628-2014 |
| Date reported | May 28, 2014 |
| Date initiated | April 4, 2014 |
| Recalling firm | Philips Medical Systems, Inc. |
| Firm location | Andover, MA |
| Affected scope | 188,520 devices |
| Distribution | Worldwide Distribution - Including US Nationwide, Canada, Czech Republic, Denmark, Netherlands, Estonia, Finland, France, Germany, Greece, Hungary, Italy, Japan, Kazakhstan, Norway, Poland, Romania, Russia, China, Slovakia, Sweden, and Tur… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 28, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.