PlainRecalls
FDA Devices Moderate Class II Ongoing

BECKMAN COULTER, INC. a. COULTER DxH Diff Pack REAGENT, DIFF PK FP DXHCOULTR, Catalog #628020. b. COULTER DxH Cleaner REAGENT, CLEANER FP DXH 10LCOULTR Catalog #628023. c. DxH 500 Series Lyse LYSE, VERSA SOL DXH 500 SERIES(0.5L/KT) Catalog # B36846. d. DxH 500 Series Cleaner CLEANER, COULTER CLENZ DXH 500SERIES (0.5L/KT) Catalog # B36868

Reported: August 31, 2022 Initiated: May 25, 2022 #Z-1629-2022

Product Description

BECKMAN COULTER, INC. a. COULTER DxH Diff Pack REAGENT, DIFF PK FP DXHCOULTR, Catalog #628020. b. COULTER DxH Cleaner REAGENT, CLEANER FP DXH 10LCOULTR Catalog #628023. c. DxH 500 Series Lyse LYSE, VERSA SOL DXH 500 SERIES(0.5L/KT) Catalog # B36846. d. DxH 500 Series Cleaner CLEANER, COULTER CLENZ DXH 500SERIES (0.5L/KT) Catalog # B36868

Reason for Recall

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Details

Units Affected
a. 11 b. 13 c. 15 d. 33
Distribution
US Nationwide
Location
Richmond, VA

Frequently Asked Questions

What product was recalled?
BECKMAN COULTER, INC. a. COULTER DxH Diff Pack REAGENT, DIFF PK FP DXHCOULTR, Catalog #628020. b. COULTER DxH Cleaner REAGENT, CLEANER FP DXH 10LCOULTR Catalog #628023. c. DxH 500 Series Lyse LYSE, VERSA SOL DXH 500 SERIES(0.5L/KT) Catalog # B36846. d. DxH 500 Series Cleaner CLEANER, COULTER CLENZ DXH 500SERIES (0.5L/KT) Catalog # B36868. Recalled by Mckesson Medical-Surgical Inc. Corporate Office. Units affected: a. 11 b. 13 c. 15 d. 33.
Why was this product recalled?
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 31, 2022. Severity: Moderate. Recall number: Z-1629-2022.