natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph

Reported: April 8, 2020 Initiated: March 5, 2020 #Z-1630-2020

Product Description

natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph

Reason for Recall

The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.

Details

Units Affected
40 units
Distribution
Worldwide distribution. US nationwide, Australia, Dominican Republic, United Arab Emirates, Israel, Italy, Hong Kong, Korea, Hungary, Switzerland, Colombia, and United Kingdom
Location
Oakville, N/A

Frequently Asked Questions

What product was recalled?
natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph. Recalled by Natus Neurology DBA Excel Tech., Ltd. (XLTEK). Units affected: 40 units.
Why was this product recalled?
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 8, 2020. Severity: Moderate. Recall number: Z-1630-2020.