PlainRecalls
FDA Devices Moderate Class II Terminated

8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202

Reported: April 8, 2020 Initiated: November 4, 2019 #Z-1635-2020

Product Description

8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202

Reason for Recall

The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.

Details

Recalling Firm
Trilliant Surgical, LLC
Units Affected
117
Distribution
Distributors in 21 states: AR, AZ, CA, FL, GA, IA,IN, KS, KY, MN, MO, NC, ND, NJ, NY, OH, TN, TX, UT, WA, and WI.
Location
Houston, TX

Frequently Asked Questions

What product was recalled?
8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202. Recalled by Trilliant Surgical, LLC. Units affected: 117.
Why was this product recalled?
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 8, 2020. Severity: Moderate. Recall number: Z-1635-2020.