Moderate
FDA Devices recall · Reported May 27, 2015
A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose…
Philips Medical Systems, Inc. recalled Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 R… - a moderate-severity action.
Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 R… was recalled by Philips Medical Systems, Inc. in May 27, 2015. Reason: A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pi…. Check the official notice for the remedy. Verify recall #Z-1643-2015 with the FDA Devices before acting.
The recall
Philips Medical Systems, Inc. issued this moderate-severity FDA Devices recall-A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pi….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1643-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1643-2015) was formally reported on May 27, 2015, with the manufacturer initiating the action on April 23, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems, Inc. is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 4.
The documented reason for this recall is: A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Spec… Distribution data in the federal record shows the product reached: US Nationwide Distribution in the states of MO, FL, OH. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4
Related Recalls
6
3 from same agency
Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data into the system, use that data to construct a plan for radiation therapy and evaluate the plan. Optionally, the qualified medical personnel may output the plan in an electronic or printed form for use by other systems in the delivery of treatment to a patient.
A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Specifically, the dose engine is being passed the wrong snout position. The snout position is used to determine the penumbra of the beam which includes the calculation of the source si
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1643-2015 |
| Date reported | May 27, 2015 |
| Date initiated | April 23, 2015 |
| Recalling firm | Philips Medical Systems, Inc. |
| Firm location | Andover, MA |
| Affected scope | 4 |
| Distribution | US Nationwide Distribution in the states of MO, FL, OH |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 27, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.