***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

Reported: July 10, 2013 Initiated: June 24, 2011 #Z-1657-2013

Product Description

***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

Reason for Recall

The RPM of the Motor Handpiece may be less than 80,000 RPM.

Details

Recalling Firm
The Anspach Effort, Inc.
Units Affected
61 units
Distribution
Worldwide Distribution - USA Nationwide including the states of: AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of: Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;
Location
Palm Beach Gardens, FL

Frequently Asked Questions

What product was recalled?
***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.. Recalled by The Anspach Effort, Inc.. Units affected: 61 units.
Why was this product recalled?
The RPM of the Motor Handpiece may be less than 80,000 RPM.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 10, 2013. Severity: Moderate. Recall number: Z-1657-2013.