FDA Devices recall · Reported May 16, 2018
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitt…
Baxter Healthcare Corporation recalled Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXR Pro… - a moderate-severity action.
Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXR Pro… was recalled by Baxter Healthcare Corporation in May 16, 2018. Reason: Service technician improperly documented final release testing on product, indicating that the full battery o…. Check the official notice for the remedy. Verify recall #Z-1665-2018 with the FDA Devices before acting.
The recall
Baxter Healthcare Corporation issued this moderate-severity FDA Devices recall-Service technician improperly documented final release testing on product, indicating that the full battery o….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1665-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1665-2018) was formally reported on May 16, 2018, with the manufacturer initiating the action on April 3, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Baxter Healthcare Corporation is listed as the recalling firm, operating out of Medina, NY. Federal records list the affected scope as 444 units in total.
The documented reason for this recall is: Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line… Distribution data in the federal record shows the product reached: US Nationwide & Puerto Rico. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
444 units in total
Related Recalls
6
3 from same agency
Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXR Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluids. Spectrum infusion pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by user. The Spectrum and Spectrum with MAster Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming.
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1665-2018 |
| Date reported | May 16, 2018 |
| Date initiated | April 3, 2018 |
| Recalling firm | Baxter Healthcare Corporation |
| Firm location | Medina, NY |
| Affected scope | 444 units in total |
| Distribution | US Nationwide & Puerto Rico |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 16, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.