PlainRecalls

CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111

Reported: May 14, 2025 Initiated: April 10, 2025 #Z-1667-2025 185,975 US; 127,701 OUS Total units

Smiths Medical ASD, Inc. issued this FDA Devices recall on May 14, 2025. Classified as Critical severity (Class I). Approximately 185,975 US; 127,701 OUS Total units are affected. The recall was issued because: Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1667-2025) was formally reported on May 14, 2025, with the manufacturer initiating the action on April 10, 2025. It is classified under Critical severity (Class I), with a current status of Ongoing. Smiths Medical ASD, Inc. is listed as the recalling firm, operating out of Minneapolis, MN. Federal records indicate 185,975 US; 127,701 OUS Total units are affected, placing this recall in the million-unit bracket that typically triggers nationwide consumer alerts and retailer sweeps.

The documented reason for this recall is: Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death. Distribution data in the federal record shows the product reached: Worldwide - US nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 1 year old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Critical

Units Affected

185,975 US; 127,701 OUS Total

Related Recalls

6

6 from same agency

Product Description

CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111-0300-09_OKA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OOMIYA 21-2111-0300-09_OMI CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OSAKA 21-2111-0300-09_OSA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR SAPPORO 21-2111-0300-09_SAP CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR TOKYO 21-2111-0300-09_TOK CADD_SOLIS_MODEL 2110_ENGLISH_1/EA 21-2111-0200-50JP CADD-Solis VIP CADD-SOLIS V4.2 LOAN FUK 21-2111-0402-09-FUK CADD-SOLIS V4.2 LOAN NAG 21-2111-0402-09-NAG CADD-SOLIS V4.2 LOAN OKA 21-2111-0402-09-OKA CADD-SOLIS V4.2 LOAN OSA 21-2111-0402-09-OSA CADD-SOLIS V4.2 LOAN SAI 21-2111-0402-09-SAI CADD-SOLIS V4.2 LOAN SAP 21-2111-0402-09-SAP CADD-SOLIS V4.2 LOAN TOKE 21-2111-0402-09-TOKE CADD-SOLIS V4.2 LOAN TOKW 21-2111-0402-09-TOKW CADD-SOLIS V4.2 LOAN YELLOW KYO 21-2112-0402-09-KYO CADD-SOLIS V4.2 LOAN YELLOW NAG 21-2112-0402-09-NAG CADD-SOLIS V4.2 LOAN YELLOW YOK ¿ 21-2112-0402-09-YOK CADD-SOLIS V4.2 LOAN YOK 21-2111-0402-09-YOK Number:21-2111-0403-02Part Type:List Number/Finished GoodLifecycle Phase:03-ReleasedDescription:PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, FRENCH, 1/EA 21-2111-0403-02 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, YELLOW, ENGLISH, 1/EA 21-2112-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA 21-2111-0300-00 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DANISH 1/EA ¿ 21-2111-0300-12 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DUTCH 1/EA ¿ 21-2111-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, FRENCH 1/EA ¿ 21-2111-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, GERMAN 1/EA ¿ 21-2111-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, ITALIAN 1/EA ¿ 21-2111-0300-07 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, JAPANESE 21-2111-0300-09JP PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, FRENCH 1/EA ¿ 21-2111-0300-232 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA ¿ 21-2111-0300-231 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, SWEDISH 1/EA 21-2111-0300-244 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 Not PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, PORTUGUESE 1/EA ¿ 21-2111-0300-17 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SPANISH 1/EA ¿ 21-2111-0300-08 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SWEDISH 1/EA ¿ 21-2111-0300-14 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, DUTCH 1/EA 21-2112-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, GERMAN 1/EA 21-2112-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA 21-2112-0300-00 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, DANISH 1/EA ¿ 21-2101-0200-12 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, FRENCH 1/EA ¿ 21-2101-0200-02 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, GERMAN 1/EA ¿ 21-2101-0200-03 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, ITALIAN 1/EA ¿ 21-2101-02

Reason for Recall

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Details

Recalling Firm
Smiths Medical ASD, Inc.
Units Affected
185,975 US; 127,701 OUS Total
Distribution
Worldwide - US nationwide distribution.
Location
Minneapolis, MN

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-1667-2025
Date reported May 14, 2025
Date initiated April 10, 2025
Recalling firm Smiths Medical ASD, Inc.
Units affected 185,975 US; 127,701 OUS Total
Distribution Worldwide - US nationwide distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

185,975 US; 127,701 OUS Total units affected — million-unit bracket.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units) ✓ This recall

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111-0300-09_OKA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OOMIYA 21-2111-0300-09_OMI CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OSAKA 21-2111-0300-09_OSA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR SAPPORO 21-2111-0300-09_SAP CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR TOKYO 21-2111-0300-09_TOK CADD_SOLIS_MODEL 2110_ENGLISH_1/EA 21-2111-0200-50JP CADD-Solis VIP CADD-SOLIS V4.2 LOAN FUK 21-2111-0402-09-FUK CADD-SOLIS V4.2 LOAN NAG 21-2111-0402-09-NAG CADD-SOLIS V4.2 LOAN OKA 21-2111-0402-09-OKA CADD-SOLIS V4.2 LOAN OSA 21-2111-0402-09-OSA CADD-SOLIS V4.2 LOAN SAI 21-2111-0402-09-SAI CADD-SOLIS V4.2 LOAN SAP 21-2111-0402-09-SAP CADD-SOLIS V4.2 LOAN TOKE 21-2111-0402-09-TOKE CADD-SOLIS V4.2 LOAN TOKW 21-2111-0402-09-TOKW CADD-SOLIS V4.2 LOAN YELLOW KYO 21-2112-0402-09-KYO CADD-SOLIS V4.2 LOAN YELLOW NAG 21-2112-0402-09-NAG CADD-SOLIS V4.2 LOAN YELLOW YOK ¿ 21-2112-0402-09-YOK CADD-SOLIS V4.2 LOAN YOK 21-2111-0402-09-YOK Number:21-2111-0403-02Part Type:List Number/Finished GoodLifecycle Phase:03-ReleasedDescription:PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, FRENCH, 1/EA 21-2111-0403-02 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, YELLOW, ENGLISH, 1/EA 21-2112-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA 21-2111-0300-00 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DANISH 1/EA ¿ 21-2111-0300-12 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DUTCH 1/EA ¿ 21-2111-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, FRENCH 1/EA ¿ 21-2111-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, GERMAN 1/EA ¿ 21-2111-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, ITALIAN 1/EA ¿ 21-2111-0300-07 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, JAPANESE 21-2111-0300-09JP PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, FRENCH 1/EA ¿ 21-2111-0300-232 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA ¿ 21-2111-0300-231 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, SWEDISH 1/EA 21-2111-0300-244 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 Not PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, PORTUGUESE 1/EA ¿ 21-2111-0300-17 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SPANISH 1/EA ¿ 21-2111-0300-08 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SWEDISH 1/EA ¿ 21-2111-0300-14 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, DUTCH 1/EA 21-2112-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, GERMAN 1/EA 21-2112-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA 21-2112-0300-00 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, DANISH 1/EA ¿ 21-2101-0200-12 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, FRENCH 1/EA ¿ 21-2101-0200-02 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, GERMAN 1/EA ¿ 21-2101-0200-03 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, ITALIAN 1/EA ¿ 21-2101-02. Recalled by Smiths Medical ASD, Inc.. Units affected: 185,975 US; 127,701 OUS Total.
Why was this product recalled?
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 14, 2025. Severity: Critical. Recall number: Z-1667-2025.
Where was the recalled product distributed?
Distribution: Worldwide - US nationwide distribution..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1667-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as critical severity, meaning the product carries a reasonable probability of serious adverse health consequences or death. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).