Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-10003-300 (SIZE 3- Right Medial); f) REF 1-10003-310 (SIZE 3- Left Medial); g) REF 1-10003-400 (SIZE 4- Right Medial); h) REF 1-10003-410 (SIZE 4- Left Medial); i) REF 1-10003-500 (SIZE 5- Right Medial); j) REF
Product Description
Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-10003-300 (SIZE 3- Right Medial); f) REF 1-10003-310 (SIZE 3- Left Medial); g) REF 1-10003-400 (SIZE 4- Right Medial); h) REF 1-10003-410 (SIZE 4- Left Medial); i) REF 1-10003-500 (SIZE 5- Right Medial); j) REF 1-10003-510 (SIZE 5- Left Medial); k) REF 1-10003-600 (SIZE 6- Right Medial); l) REF 1-10003-610 (SIZE 6- Left Medial); m) REF 1-10003-700 (SIZE 7- Right Medial); n) REF 1-10003-710 (SIZE 7- Left Medial); o) REF 1-10003-800, (SIZE 8 - Right Medial); p) REF 1-10003-810, (SIZE 8 - Left Medial); Unicondylar knee prothesis
Reason for Recall
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Details
- Recalling Firm
- Smith & Nephew, Inc.
- Units Affected
- 1994 units
- Distribution
- US Nationwide distribution.
- Location
- Memphis, TN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11