PlainRecalls
FDA Devices Moderate Class II Terminated

GS777 Wall Transformer Model: 77710 Intended Use: The GS777 Wall Transformer is intended for use in ambulatory and acute care settings.

Reported: June 2, 2021 Initiated: April 22, 2021 #Z-1672-2021

Product Description

GS777 Wall Transformer Model: 77710 Intended Use: The GS777 Wall Transformer is intended for use in ambulatory and acute care settings.

Reason for Recall

Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock

Details

Recalling Firm
Welch Allyn Inc
Units Affected
789682
Distribution
Worldwide - US Nationwide distribution Foreign: Afghanistan Albania Algeria American Samoa Argentina Australia Austria Azerbaijan Bahamas (the) Bahrain Bangladesh Barbados Belgium Bermuda Bhutan Bolivia (Plurinational State of) Botswana Brazil Brunei Darussalam Bulgaria Cameroon Canada Chad Chile China Colombia Congo (the Democratic Republic of the) Congo (the) Costa Rica Croatia Cyprus Czechia Denmark Dominican Republic (the) Ecuador Egypt El Salvador Estonia Eswatini Finland France French Polynesia Georgia Germany Ghana Greece Guam Guatemala Honduras Hong Kong Hungary Iceland India Indonesia Iran (Islamic Republic of) Iraq Ireland Israel Italy Jamaica Japan Jordan Kenya Korea (the Republic of) Kuwait Lao People's Democratic Republic (the) Latvia Lebanon Lesotho Libya Luxembourg Malaysia Maldives Malta Mexico Mozambique Myanmar Namibia Nepal Netherlands (the) New Zealand Nicaragua Nigeria Northern Mariana Islands (the) Norway Oman Pakistan Panama Paraguay Peru Philippines (the) Poland Portugal Puerto Rico Qatar R¿union Romania Russian Federation (the) Saudi Arabia Senegal Serbia Singapore Slovenia South Africa Spain Sri Lanka Sweden Switzerland Taiwan (Province of China) Thailand Trinidad and Tobago Tunisia Turkey Uganda Ukraine United Arab Emirates (the) United Kingdom Venezuela (Bolivarian Republic of) Viet Nam
Location
Skaneateles Falls, NY

Frequently Asked Questions

What product was recalled?
GS777 Wall Transformer Model: 77710 Intended Use: The GS777 Wall Transformer is intended for use in ambulatory and acute care settings.. Recalled by Welch Allyn Inc. Units affected: 789682.
Why was this product recalled?
Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock
Which agency issued this recall?
This recall was issued by the FDA Devices on June 2, 2021. Severity: Moderate. Recall number: Z-1672-2021.