DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724

Reported: June 14, 2023 Initiated: April 13, 2023 #Z-1677-2023

Product Description

DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724

Reason for Recall

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Details

Units Affected
805 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
Location
Norwood, MA

Frequently Asked Questions

What product was recalled?
DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724. Recalled by DePuy Mitek, Inc., a Johnson & Johnson Co.. Units affected: 805 units.
Why was this product recalled?
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Which agency issued this recall?
This recall was issued by the FDA Devices on June 14, 2023. Severity: Moderate. Recall number: Z-1677-2023.