Severity
Moderate
FDA Devices recall · Reported May 7, 2025
There is a leakage issue associated with the tubing in the manifold of the instrument.
Leica Biosystems Melbourne Pty Ltd recalled HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4… - a moderate-severity action.
HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4… was recalled by Leica Biosystems Melbourne Pty Ltd in May 7, 2025. Reason: There is a leakage issue associated with the tubing in the manifold of the instrument.. Check the official notice for the remedy. Verify recall #Z-1683-2025 with the FDA Devices before acting.
The recall
Leica Biosystems Melbourne Pty Ltd issued this moderate-severity FDA Devices recall-There is a leakage issue associated with the tubing in the manifold of the instrument..
Sourced from official FDA Devices enforcement records. Verify recall #Z-1683-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1683-2025) was formally reported on May 7, 2025, with the manufacturer initiating the action on April 2, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Leica Biosystems Melbourne Pty Ltd is listed as the recalling firm, operating out of Mount Waverley. Federal records list the affected scope as 77 units.
The documented reason for this recall is: There is a leakage issue associated with the tubing in the manifold of the instrument. Distribution data in the federal record shows the product reached: US States: AZ, CA, FL, MN, NJ, NY, NC, OH, PA, TN, TX, UT, VA, and WI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
77 units
Related Recalls
6
3 from same agency
HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.
There is a leakage issue associated with the tubing in the manifold of the instrument.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1683-2025 |
| Date reported | May 7, 2025 |
| Date initiated | April 2, 2025 |
| Recalling firm | Leica Biosystems Melbourne Pty Ltd |
| Firm location | Mount Waverley |
| Affected scope | 77 units |
| Distribution | US States: AZ, CA, FL, MN, NJ, NY, NC, OH, PA, TN, TX, UT, VA, and WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 7, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.