PlainRecalls
FDA Devices Low Class III Ongoing

RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026578

Reported: June 14, 2023 Initiated: April 14, 2023 #Z-1695-2023

Product Description

RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026578

Reason for Recall

QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.

Details

Recalling Firm
Mechatronics USA
Units Affected
10752 units
Distribution
Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.
Location
Warwick, RI

Frequently Asked Questions

What product was recalled?
RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026578. Recalled by Mechatronics USA. Units affected: 10752 units.
Why was this product recalled?
QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 14, 2023. Severity: Low. Recall number: Z-1695-2023.