CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS

Reported: June 12, 2019 Initiated: April 10, 2018 #Z-1696-2019

Product Description

CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS

Reason for Recall

There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.

Details

Recalling Firm
Accuray Incorporated
Units Affected
68 units
Distribution
US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
Location
Madison, WI

Frequently Asked Questions

What product was recalled?
CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS. Recalled by Accuray Incorporated. Units affected: 68 units.
Why was this product recalled?
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 12, 2019. Severity: Moderate. Recall number: Z-1696-2019.