BEQ-TOP 40700 HUNTINGTON, Catalog No. 701062899
Reported: May 16, 2018 Initiated: October 19, 2017 #Z-1711-2018
Product Description
BEQ-TOP 40700 HUNTINGTON, Catalog No. 701062899
Reason for Recall
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Details
- Recalling Firm
- Datascope Corporation
- Units Affected
- 4
- Distribution
- Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
- Location
- Fairfield, NJ
Frequently Asked Questions
What product was recalled? ▼
BEQ-TOP 40700 HUNTINGTON, Catalog No. 701062899. Recalled by Datascope Corporation. Units affected: 4.
Why was this product recalled? ▼
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 16, 2018. Severity: Moderate. Recall number: Z-1711-2018.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11