Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Reported: April 22, 2020 Initiated: January 21, 2020 #Z-1717-2020

Product Description

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Reason for Recall

When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.

Details

Recalling Firm
Sysmex America, Inc.
Units Affected
40 units
Distribution
USA: FL, VT, IL.
Location
Mundelein, IL