Washer Disinfector Aquadis 56; Model Number: 56M

Reported: May 14, 2025 Initiated: April 2, 2025 #Z-1721-2025

Product Description

Washer Disinfector Aquadis 56; Model Number: 56M

Reason for Recall

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Details

Recalling Firm
Getinge Disinfection Ab
Units Affected
424 units (79 US, 345 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.
Location
Vaxjo, N/A

Frequently Asked Questions

What product was recalled?
Washer Disinfector Aquadis 56; Model Number: 56M. Recalled by Getinge Disinfection Ab. Units affected: 424 units (79 US, 345 OUS).
Why was this product recalled?
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.