Washer Disinfector Aquadis 56; Model Number: 56M
Reported: May 14, 2025 Initiated: April 2, 2025 #Z-1721-2025
Product Description
Washer Disinfector Aquadis 56; Model Number: 56M
Reason for Recall
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Details
- Recalling Firm
- Getinge Disinfection Ab
- Units Affected
- 424 units (79 US, 345 OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.
- Location
- Vaxjo, N/A
Frequently Asked Questions
What product was recalled? ▼
Washer Disinfector Aquadis 56; Model Number: 56M. Recalled by Getinge Disinfection Ab. Units affected: 424 units (79 US, 345 OUS).
Why was this product recalled? ▼
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 14, 2025. Severity: Moderate. Recall number: Z-1721-2025.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11