FDA Devices recall · Reported June 18, 2014

enGen (TM) Laboratory Automation System Configured with TCAutomation" Software v3.5 -- For in vitro quantitative measurement of a variety of analytes of clinical interest --- The VITROS 5,1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Che

Software anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Systems Configured with TCAutomation (TM) Software v3.5.

Recall #
Z-1741-2014
Affected scope
Domestic: 6 units, Foreign: 3 units
Initiated
April 17, 2014
Verify with FDA Devices →
View my saved recalls

Ortho-Clinical Diagnostics recalled enGen (TM) Laboratory Automation System Configured with TCAutomation" Software v3.5 -- Fo… - a moderate-severity action.

enGen (TM) Laboratory Automation System Configured with TCAutomation" Software v3.5 -- Fo… was recalled by Ortho-Clinical Diagnostics in June 18, 2014. Reason: Software anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Systems Configu…. Check the official notice for the remedy. Verify recall #Z-1741-2014 with the FDA Devices before acting.

The recall

Ortho-Clinical Diagnostics issued this moderate-severity FDA Devices recall-Software anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Systems Configu….

Moderate
severity level
Class II
classification
June 18, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1741-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

enGen (TM) Laboratory Automation System Configured with TCAutomation" Software v3.5 -- For in vitro quantitative measurement of a variety of analytes of clinical interest --- The VITROS 5,1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.). ---The enGen (TM) Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TCAutomation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen (TM) System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. By using the enGen (TM) Laboratory Automation System as a communication interface between the LIS and the clinical analyzers in the laboratory, the user can manage the workflow to the analyzers and other devices as an integrated system. This eliminates the need for the user to program tests for each sample uniquely on each analyzer. Within the enGen (TM) Laboratory Automation System, the physical routing of samples to the correct location is also automatic. The Buffer Module functions either as a location for samples waiting for results or as a storage area for empty sample tube carriers. When a sample is presented to a clinical analyzer via the Bypass module and is metered for testing, the sample remains on the track system in the Buffer module while waiting for the test results to be generated. If all test results are generated as expected and no additional actions are required, the sample is routed to a designated Exit module where it can be removed from the track system by the user. However, there are certain situations in which a follow-up action (retest of the sample) may be necessary. In these situations, the sample is routed back to a clinical analyzer via the Bypass module to complete the follow-up action. Examples of situations requiring a retest include, but are not limited to: repeat testing when a "No Result"is generated during the initial test event, sample dilution when an out of range result is obtained during the initial test event, and routing to an alternate analyzer when the first analyzer the sample is sent to is unavailable for testing. Once the follow-up sample metering actions are complete, the sample is routed to Buffer module where it remains until testing is complete. Once all test results are obtained the sample is routed to the designated Exit module.

Reason for recall

Software anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Systems Configured with TCAutomation (TM) Software v3.5.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1741-2014
Date reported June 18, 2014
Date initiated April 17, 2014
Recalling firm Ortho-Clinical Diagnostics
Firm location Rochester, NY
Affected scope Domestic: 6 units, Foreign: 3 units
Distribution Worldwide Distribution - US including the states of Arizona, Delaware, Ohio, Tennessee, and North Carolina; and the countries of Mexico, England and Italy.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1741-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
enGen (TM) Laboratory Automation System Configured with TCAutomation" Software v3.5 -- For in vitro quantitative measurement of a variety of analytes of clinical interest --- The VITROS 5,1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.). ---The enGen (TM) Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TCAutomation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen (TM) System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. By using the enGen (TM) Laboratory Automation System as a communication interface between the LIS and the clinical analyzers in the laboratory, the user can manage the workflow to the analyzers and other devices as an integrated system. This eliminates the need for the user to program tests for each sample uniquely on each analyzer. Within the enGen (TM) Laboratory Automation System, the physical routing of samples to the correct location is also automatic. The Buffer Module functions either as a location for samples waiting for results or as a storage area for empty sample tube carriers. When a sample is presented to a clinical analyzer via the Bypass module and is metered for testing, the sample remains on the track system in the Buffer module while waiting for the test results to be generated. If all test results are generated as expected and no additional actions are required, the sample is routed to a designated Exit module where it can be removed from the track system by the user. However, there are certain situations in which a follow-up action (retest of the sample) may be necessary. In these situations, the sample is routed back to a clinical analyzer via the Bypass module to complete the follow-up action. Examples of situations requiring a retest include, but are not limited to: repeat testing when a "No Result"is generated during the initial test event, sample dilution when an out of range result is obtained during the initial test event, and routing to an alternate analyzer when the first analyzer the sample is sent to is unavailable for testing. Once the follow-up sample metering actions are complete, the sample is routed to Buffer module where it remains until testing is complete. Once all test results are obtained the sample is routed to the designated Exit module.. Recalled by Ortho-Clinical Diagnostics. Units affected: Domestic: 6 units, Foreign: 3 units.
Why was this product recalled?
Software anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Systems Configured with TCAutomation (TM) Software v3.5.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 18, 2014. Severity: Moderate. Recall number: Z-1741-2014.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US including the states of Arizona, Delaware, Ohio, Tennessee, and North Carolina; and the countries of Mexico, England and Italy..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported June 18, 2014. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026