RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This d

Reported: July 24, 2013 Initiated: July 1, 2013 #Z-1742-2013

Product Description

RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM).

Reason for Recall

The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

Details

Units Affected
9
Distribution
Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.
Location
Portage, MI

Frequently Asked Questions

What product was recalled?
RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM).. Recalled by Stryker Instruments Div. of Stryker Corporation. Units affected: 9.
Why was this product recalled?
The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100
Which agency issued this recall?
This recall was issued by the FDA Devices on July 24, 2013. Severity: Moderate. Recall number: Z-1742-2013.