The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Reported: September 28, 2022 Initiated: August 1, 2022 #Z-1772-2022

Product Description

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Reason for Recall

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Details

Recalling Firm
Randox Laboratories Ltd.
Units Affected
2 units
Distribution
US Nationwide distribution in the states of WV, NC.
Location
Crumlin (North), N/A

Frequently Asked Questions

What product was recalled?
The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.. Recalled by Randox Laboratories Ltd.. Units affected: 2 units.
Why was this product recalled?
***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg
Which agency issued this recall?
This recall was issued by the FDA Devices on September 28, 2022. Severity: Moderate. Recall number: Z-1772-2022.