Medivators Endo Smartcap Endoscope Tubing. REF100145CO2EXT, packaged as Contents 1, 10, 50, and found in BMP-C145CO2EXT. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 Series Endoscopes With Extended CO2 Input. Sterile EO - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.
Product Description
Medivators Endo Smartcap Endoscope Tubing. REF100145CO2EXT, packaged as Contents 1, 10, 50, and found in BMP-C145CO2EXT. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 Series Endoscopes With Extended CO2 Input. Sterile EO - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.
Reason for Recall
Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
Details
- Recalling Firm
- Medivators, Inc.
- Units Affected
- 53,790 units total
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Australia, England.
- Location
- Plymouth, MN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11