Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

Reported: June 19, 2019 Initiated: May 20, 2019 #Z-1778-2019

Product Description

Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

Reason for Recall

Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.

Details

Units Affected
40 units
Distribution
Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine
Location
Wayne, NJ

Frequently Asked Questions

What product was recalled?
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.. Recalled by Maquet Cardiovascular Us Sales, Llc. Units affected: 40 units.
Why was this product recalled?
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 19, 2019. Severity: Moderate. Recall number: Z-1778-2019.