Severity
Moderate
FDA Devices recall · Reported April 19, 2017
GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time …
Ge Healthcare, LLC recalled Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD. - a moderate-severity action.
Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD. was recalled by Ge Healthcare, LLC in April 19, 2017. Reason: GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by…. Check the official notice for the remedy. Verify recall #Z-1780-2017 with the FDA Devices before acting.
The recall
Ge Healthcare, LLC issued this moderate-severity FDA Devices recall-GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1780-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1780-2017) was formally reported on April 19, 2017, with the manufacturer initiating the action on March 10, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Ge Healthcare, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records list the affected scope as 396 (67 US; 329 OUS).
The documented reason for this recall is: GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certai… Distribution data in the federal record shows the product reached: United Kingdom, United Arab Emirates, Turkey, Thailand, Switzerland, Sweden, Spain, South Africa, Slovenia, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Norway, New Zealand, New Caledonia, Netherlands…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
396 (67 US; 329 OUS)
Related Recalls
6
3 from same agency
Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.
GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1780-2017 |
| Date reported | April 19, 2017 |
| Date initiated | March 10, 2017 |
| Recalling firm | Ge Healthcare, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 396 (67 US; 329 OUS) |
| Distribution | United Kingdom, United Arab Emirates, Turkey, Thailand, Switzerland, Sweden, Spain, South Africa, Slovenia, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Norway, New Zealand, New Caledonia, Netherlands, Myanmar, Kuwait,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 19, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.