Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Reported: April 29, 2020 Initiated: March 2, 2020 #Z-1790-2020

Product Description

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Reason for Recall

Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

Details

Recalling Firm
CSL Behring GmbH
Units Affected
297 PCK US; 1247 PCK OUS
Distribution
IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan
Location
Marburg, N/A

Frequently Asked Questions

What product was recalled?
Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997. Recalled by CSL Behring GmbH. Units affected: 297 PCK US; 1247 PCK OUS.
Why was this product recalled?
Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring
Which agency issued this recall?
This recall was issued by the FDA Devices on April 29, 2020. Severity: Moderate. Recall number: Z-1790-2020.