Severity
Moderate
FDA Devices recall · Reported June 16, 2021
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.…
Philips North America Llc recalled Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 202… - a moderate-severity action.
Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 202… was recalled by Philips North America Llc in June 16, 2021. Reason: The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are …. Check the official notice for the remedy. Verify recall #Z-1795-2021 with the FDA Devices before acting.
The recall
Philips North America Llc issued this moderate-severity FDA Devices recall-The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1795-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1795-2021) was formally reported on June 16, 2021, with the manufacturer initiating the action on April 23, 2021. It is classified under Moderate severity (Class II), with a current status of Ongoing. Philips North America Llc is listed as the recalling firm, operating out of Cambridge, MA. Federal records list the affected scope as 156 MR Systems.
The documented reason for this recall is: The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving acti… Distribution data in the federal record shows the product reached: Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerlan…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
156 MR Systems
Related Recalls
6
3 from same agency
Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109 - Ingenia Ambition X
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1795-2021 |
| Date reported | June 16, 2021 |
| Date initiated | April 23, 2021 |
| Recalling firm | Philips North America Llc |
| Firm location | Cambridge, MA |
| Affected scope | 156 MR Systems |
| Distribution | Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 16, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.