PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

Reported: May 6, 2020 Initiated: March 11, 2020 #Z-1796-2020

Product Description

PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

Reason for Recall

During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.

Details

Units Affected
1932
Distribution
U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
Location
South Jordan, UT

Frequently Asked Questions

What product was recalled?
PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT. Recalled by Merit Medical Systems, Inc.. Units affected: 1932.
Why was this product recalled?
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 6, 2020. Severity: Moderate. Recall number: Z-1796-2020.