PlainRecalls
FDA Devices Moderate Class II Terminated

BD BBL BCYE Selective Agar with PAV, Catalog # 297880, packaged in cartons of 10 plates, and labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 USA 800-638-8663 www.bd.com/ds*** BCYE Selective with PAV is used in qualitative procedures for isolation of Legionella species from clinical specimen and nonclinical (environmental) samples. It is similar to the Edelstein formula, except that the concentration of polymyxin B is reduced by half, and vancomycin is subst

Reported: July 31, 2013 Initiated: June 6, 2013 #Z-1804-2013

Product Description

BD BBL BCYE Selective Agar with PAV, Catalog # 297880, packaged in cartons of 10 plates, and labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 USA 800-638-8663 www.bd.com/ds*** BCYE Selective with PAV is used in qualitative procedures for isolation of Legionella species from clinical specimen and nonclinical (environmental) samples. It is similar to the Edelstein formula, except that the concentration of polymyxin B is reduced by half, and vancomycin is substituted for cefamandole.

Reason for Recall

Microbiological identification media may exhibit reduced levels of Vancomycin.

Details

Recalling Firm
Becton Dickinson & Co.
Units Affected
163 cartons
Distribution
Worldwide distribution: US (nationwide) and countries of:Belgium, Canada, Columbia, Mexico, Singapore, and Taiwan.
Location
Sparks, MD

Frequently Asked Questions

What product was recalled?
BD BBL BCYE Selective Agar with PAV, Catalog # 297880, packaged in cartons of 10 plates, and labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 USA 800-638-8663 www.bd.com/ds*** BCYE Selective with PAV is used in qualitative procedures for isolation of Legionella species from clinical specimen and nonclinical (environmental) samples. It is similar to the Edelstein formula, except that the concentration of polymyxin B is reduced by half, and vancomycin is substituted for cefamandole.. Recalled by Becton Dickinson & Co.. Units affected: 163 cartons.
Why was this product recalled?
Microbiological identification media may exhibit reduced levels of Vancomycin.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 31, 2013. Severity: Moderate. Recall number: Z-1804-2013.