Severity
Moderate
FDA Devices recall · Reported June 25, 2014
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can b…
Raysearch Laboratories Ab recalled RayStation software version 4.0.3.4 RayStation is a software system designed for treat… - a moderate-severity action.
RayStation software version 4.0.3.4 RayStation is a software system designed for treat… was recalled by Raysearch Laboratories Ab in June 25, 2014. Reason: This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable …. Check the official notice for the remedy. Verify recall #Z-1806-2014 with the FDA Devices before acting.
The recall
Raysearch Laboratories Ab issued this moderate-severity FDA Devices recall-This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1806-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1806-2014) was formally reported on June 25, 2014, with the manufacturer initiating the action on April 25, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Raysearch Laboratories Ab is listed as the recalling firm, operating out of Stockholm, N/A. Federal records list the affected scope as 25 sites/245 units (US), 25 sites/124 units (foreign).
The documented reason for this recall is: This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume pr… Distribution data in the federal record shows the product reached: Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
25 sites/245 units (US), 25 sites/124 units (foreign)
Related Recalls
6
3 from same agency
RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1806-2014 |
| Date reported | June 25, 2014 |
| Date initiated | April 25, 2014 |
| Recalling firm | Raysearch Laboratories Ab |
| Firm location | Stockholm, N/A |
| Affected scope | 25 sites/245 units (US), 25 sites/124 units (foreign) |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerla… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 25, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.