Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replaceme…
Smith & Nephew, Inc. recalled META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690 - a moderate-severity action.
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690 was recalled by Smith & Nephew, Inc. in May 28, 2025. Reason: Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, whic…. Check the official notice for the remedy. Verify recall #Z-1806-2025 with the FDA Devices before acting.
The recall
Smith & Nephew, Inc. issued this moderate-severity FDA Devices recall-Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, whic….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1806-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1806-2025) was formally reported on May 28, 2025, with the manufacturer initiating the action on March 18, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Smith & Nephew, Inc. is listed as the recalling firm, operating out of Memphis, TN. Federal records list the affected scope as 75.
The documented reason for this recall is: Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
75
Related Recalls
6
3 from same agency
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1806-2025 |
| Date reported | May 28, 2025 |
| Date initiated | March 18, 2025 |
| Recalling firm | Smith & Nephew, Inc. |
| Firm location | Memphis, TN |
| Affected scope | 75 |
| Distribution | Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 28, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.