Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.
Reported: May 6, 2020 Initiated: March 18, 2020 #Z-1810-2020
Product Description
Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.
Reason for Recall
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Details
- Recalling Firm
- WELCH ALLYN, INC/MORTARA
- Units Affected
- 78,521 units total
- Distribution
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
- Location
- Milwaukee, WI
Frequently Asked Questions
What product was recalled? ▼
Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.. Recalled by WELCH ALLYN, INC/MORTARA. Units affected: 78,521 units total.
Why was this product recalled? ▼
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 6, 2020. Severity: Moderate. Recall number: Z-1810-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11