Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.

Reported: May 6, 2020 Initiated: March 18, 2020 #Z-1810-2020

Product Description

Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.

Reason for Recall

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Details

Recalling Firm
WELCH ALLYN, INC/MORTARA
Units Affected
78,521 units total
Distribution
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Location
Milwaukee, WI

Frequently Asked Questions

What product was recalled?
Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.. Recalled by WELCH ALLYN, INC/MORTARA. Units affected: 78,521 units total.
Why was this product recalled?
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 6, 2020. Severity: Moderate. Recall number: Z-1810-2020.