Severity
Moderate
FDA Devices recall · Reported August 7, 2013
The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use (RTU) bottles from other suppliers than Leica. In case of a pressure …
Leica Microsystems, Inc. recalled Leica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor … - a moderate-severity action.
Leica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor … was recalled by Leica Microsystems, Inc. in August 7, 2013. Reason: The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use…. Check the official notice for the remedy. Verify recall #Z-1815-2013 with the FDA Devices before acting.
The recall
Leica Microsystems, Inc. issued this moderate-severity FDA Devices recall-The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1815-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1815-2013) was formally reported on August 7, 2013, with the manufacturer initiating the action on November 13, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Leica Microsystems, Inc. is listed as the recalling firm, operating out of Buffalo Grove, IL. Federal records list the affected scope as 128 machines.
The documented reason for this recall is: The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use (RTU) bottles from other suppliers than Leica. In case of a pressure failure of the device in combination wit… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and internationally to Australia, Belgium, Brazil, Chile, France, Germany, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, Philipines, Singapore, South Africa, Spain, Swi…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
128 machines
Related Recalls
6
3 from same agency
Leica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor for the following laboratory applications: fixation, dehydration, infiltration with intermedium, and the paraffin infiltration of histological tissue specimens.
The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use (RTU) bottles from other suppliers than Leica. In case of a pressure failure of the device in combination with a bottle with incorrect dimensions, overflow can cause contamination of reagents. Additionally, the current instructions for use does not highlight the importance to check the filling levels of all reagent bottles (RTU and system bottles) prior to every process run. Without this important information, risk of tissue damage or loss is possible. Therefore the missing information will be distributed to all customers by a Field Service Notice.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1815-2013 |
| Date reported | August 7, 2013 |
| Date initiated | November 13, 2012 |
| Recalling firm | Leica Microsystems, Inc. |
| Firm location | Buffalo Grove, IL |
| Affected scope | 128 machines |
| Distribution | Worldwide Distribution - USA (nationwide) and internationally to Australia, Belgium, Brazil, Chile, France, Germany, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, Philipines, Singapore, South Africa, Spain, Switzerland, Thailand, … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 7, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.