FDA Devices recall · Reported April 26, 2017
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R …
Philips Electronics North America Corporation recalled UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Ta… — a moderate-severity action.
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Ta… was recalled by Philips Electronics North America Corporation in April 26, 2017. Reason: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. I…. Check the official notice for the remedy. Verify recall #Z-1821-2017 with the FDA Devices before acting.
The recall
Philips Electronics North America Corporation issued this moderate-severity FDA Devices recall — Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. I….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1821-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1821-2017) was formally reported on April 26, 2017, with the manufacturer initiating the action on March 22, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Electronics North America Corporation is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 6992 (total for all devices).
The documented reason for this recall is: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, w… Distribution data in the federal record shows the product reached: Nationwide, worldwide including Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category — 40,409 medical devices recalls on record
Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
6992 (total for all devices)
Related Recalls
6
6 from same agency
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1821-2017 |
| Date reported | April 26, 2017 |
| Date initiated | March 22, 2017 |
| Recalling firm | Philips Electronics North America Corporation |
| Firm location | Andover, MA |
| Affected scope | 6992 (total for all devices) |
| Distribution | Nationwide, worldwide including Canada |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Keep tracking product safety across the federal recall archive.
Recall Checker
Search the full archive by product name, brand, or recall number across every agency.
Check a product →
FDA Devices recalls
Every recall issued by FDA Devices, newest first.
Browse the feed →
RecallRadar
Live feed of the latest recalls across the FDA, CPSC and NHTSA — filter by agency and severity.
View the live feed →
Rankings
The largest recalls by units affected and the most-recalled product categories.
See the rankings →
Browse by category
Find recalls by product type to spot recurring defect patterns.
All categories →
What to do next
A step-by-step guide to refunds, repairs, and returns after a recall.
Read the guide →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Medline Industries, LP · 2026-06-03
Abiomed, Inc. · 2026-06-03
GE Medical Systems Information Technologies Inc · 2026-06-03
Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →
Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported April 26, 2017.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.