ModerateClass IIOngoing

FDA Devices recall · Reported April 22, 2026

Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patie

Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl…

Recall #
Z-1822-2026
Affected scope
5826
Initiated
May 16, 2025
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Therakos Development Limited recalled Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Ca… - a moderate-severity action.

Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Ca… was recalled by Therakos Development Limited in April 22, 2026. Reason: Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation…. Check the official notice for the remedy. Verify recall #Z-1822-2026 with the FDA Devices before acting.

The recall

Therakos Development Limited issued this moderate-severity FDA Devices recall-Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation….

Moderate
severity level
6K units
affected scope
Class II
classification
April 22, 2026
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1822-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1822-2026) was formally reported on April 22, 2026, with the manufacturer initiating the action on May 16, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Therakos Development Limited is listed as the recalling firm, operating out of Blanchardstown, N/A. Federal records list the affected scope as 5826.

The documented reason for this recall is: Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment pha… Distribution data in the federal record shows the product reached: US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

5826

Related Recalls

6

3 from same agency

Product description

Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable

Reason for recall

Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1822-2026
Date reported April 22, 2026
Date initiated May 16, 2025
Recalling firm Therakos Development Limited
Firm location Blanchardstown, N/A
Affected scope 5826
Distribution US Nationwide distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

5826 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1822-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable. Recalled by Therakos Development Limited. Units affected: 5826.
Why was this product recalled?
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 22, 2026. Severity: Moderate. Recall number: Z-1822-2026.
Where was the recalled product distributed?
Distribution: US Nationwide distribution..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1822-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 22, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.