Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.
Reported: May 22, 2024 Initiated: April 17, 2024 #Z-1824-2024
Product Description
Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.
Reason for Recall
The guide tip can become damaged and result in the tip detaching.
Details
- Recalling Firm
- LeMaitre Vascular, Inc.
- Units Affected
- 5,604 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, and Spain.
- Location
- Burlington, MA
Frequently Asked Questions
What product was recalled? ▼
Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.. Recalled by LeMaitre Vascular, Inc.. Units affected: 5,604 units.
Why was this product recalled? ▼
The guide tip can become damaged and result in the tip detaching.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 22, 2024. Severity: Moderate. Recall number: Z-1824-2024.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11