PlainRecalls
FDA Devices Critical Class I Ongoing

Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170060

Reported: July 5, 2023 Initiated: May 25, 2023 #Z-1829-2023

Product Description

Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170060

Reason for Recall

Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.

Details

Recalling Firm
TELEFLEX LLC
Units Affected
9400 units
Distribution
US Nationwide distribution including Puerto Rico.
Location
Morrisville, NC

Frequently Asked Questions

What product was recalled?
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170060. Recalled by TELEFLEX LLC. Units affected: 9400 units.
Why was this product recalled?
Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 5, 2023. Severity: Critical. Recall number: Z-1829-2023.