RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .

Reported: June 16, 2021 Initiated: April 21, 2021 #Z-1838-2021

Product Description

RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .

Reason for Recall

Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.

Details

Recalling Firm
RAYSEARCH LABORATORIES AB
Units Affected
Domestic: 13,722 systems
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AZ, CA, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, Colombia, Czech Republic, Denmark, Egypt, Estonia, France, French Polynesia, Germany, Hong Kong, China, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK,
Location
Stockholm, N/A

Frequently Asked Questions

What product was recalled?
RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .. Recalled by RAYSEARCH LABORATORIES AB. Units affected: Domestic: 13,722 systems.
Why was this product recalled?
Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 16, 2021. Severity: Moderate. Recall number: Z-1838-2021.