Severity
Moderate
FDA Devices recall · Reported May 13, 2020
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable t…
Philips Ultrasound Inc recalled Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory … - a moderate-severity action.
Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory … was recalled by Philips Ultrasound Inc in May 13, 2020. Reason: The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properl…. Check the official notice for the remedy. Verify recall #Z-1843-2020 with the FDA Devices before acting.
The recall
Philips Ultrasound Inc issued this moderate-severity FDA Devices recall-The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properl….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1843-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1843-2020) was formally reported on May 13, 2020, with the manufacturer initiating the action on August 30, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Ultrasound Inc is listed as the recalling firm, operating out of Bothell, WA. Federal records list the affected scope as N/A.
The documented reason for this recall is: The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on t… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
N/A
Related Recalls
6
3 from same agency
Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1843-2020 |
| Date reported | May 13, 2020 |
| Date initiated | August 30, 2019 |
| Recalling firm | Philips Ultrasound Inc |
| Firm location | Bothell, WA |
| Affected scope | N/A |
| Distribution | Worldwide distribution - US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 13, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.