EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table

Reported: June 8, 2016 Initiated: April 6, 2016 #Z-1844-2016

Product Description

EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table

Reason for Recall

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.

Details

Units Affected
458
Distribution
Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.
Location
Grand Rapids, MI

Frequently Asked Questions

What product was recalled?
EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table. Recalled by Skytron, Div. The KMW Group, Inc. Units affected: 458.
Why was this product recalled?
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 8, 2016. Severity: Moderate. Recall number: Z-1844-2016.