Ingenia Ambition S Model Number (REF): (1) 782139; (2) 782133; (3) 782108; (4) 781359

Reported: May 29, 2024 Initiated: May 3, 2024 #Z-1850-2024

Product Description

Ingenia Ambition S Model Number (REF): (1) 782139; (2) 782133; (3) 782108; (4) 781359

Reason for Recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Details

Recalling Firm
Philips North America Llc
Units Affected
328 units
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Location
Cambridge, MA