PlainRecalls
FDA Devices Moderate Class II Ongoing

Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system

Reported: June 4, 2025 Initiated: March 31, 2025 #Z-1859-2025

Product Description

Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system

Reason for Recall

During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.

Details

Recalling Firm
Ultra Clean Systems, Inc.
Units Affected
96 systems
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AZ, CA, CO, FL, ID, IL, IN, KS, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SD, TX, VA, WA, WI, and WY. The country of Canada.
Location
Oldsmar, FL

Frequently Asked Questions

What product was recalled?
Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system. Recalled by Ultra Clean Systems, Inc.. Units affected: 96 systems.
Why was this product recalled?
During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 4, 2025. Severity: Moderate. Recall number: Z-1859-2025.