2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.

Reported: May 29, 2024 Initiated: April 17, 2024 #Z-1869-2024

Product Description

2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.

Reason for Recall

An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

Details

Units Affected
334 units
Distribution
US Nationwide distribution.
Location
Waltham, MA

Frequently Asked Questions

What product was recalled?
2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.. Recalled by Fresenius Medical Care Holdings, Inc.. Units affected: 334 units.
Why was this product recalled?
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 29, 2024. Severity: Moderate. Recall number: Z-1869-2024.