FDA Devices Low Class III Terminated

Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE eq

Reported: July 3, 2019 Initiated: May 10, 2019 #Z-1877-2019

Product Description

Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Reason for Recall

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Details

Recalling Firm
GE Healthcare, LLC
Units Affected
58,045 units (31,108 US; 26937 OUS) in total
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Location
Waukesha, WI

Frequently Asked Questions

What product was recalled?
Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.. Recalled by GE Healthcare, LLC. Units affected: 58,045 units (31,108 US; 26937 OUS) in total.
Why was this product recalled?
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 3, 2019. Severity: Low. Recall number: Z-1877-2019.