FDA Devices recall · Reported May 29, 2024
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of t…
Fujifilm Healthcare Americas Corporation recalled Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24 - a moderate-severity action.
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24 was recalled by Fujifilm Healthcare Americas Corporation in May 29, 2024. Reason: AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is …. Check the official notice for the remedy. Verify recall #Z-1884-2024 with the FDA Devices before acting.
The recall
Fujifilm Healthcare Americas Corporation issued this moderate-severity FDA Devices recall-AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is ….
- Moderate
- severity level
- 418 units
- affected scope
- Class II
- classification
- May 29, 2024
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1884-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
Reason for recall
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1884-2024 |
| Date reported | May 29, 2024 |
| Date initiated | April 11, 2024 |
| Recalling firm | Fujifilm Healthcare Americas Corporation |
| Firm location | Lexington, MA |
| Affected scope | 418 units |
| Distribution | US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1884-2024) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026