Probe Cover, Lite poly, 6"x48", w two bands- Ultrasonic Transducer Cover Item ID: A-AU-1002N
Reported: June 30, 2021 Initiated: April 29, 2021 #Z-1913-2021
Product Description
Probe Cover, Lite poly, 6"x48", w two bands- Ultrasonic Transducer Cover Item ID: A-AU-1002N
Reason for Recall
No 510k for the product to be used in a natural or surgical opening to the body
Details
- Recalling Firm
- Exact Medical Manufacturing, Inc.
- Units Affected
- 5621 units
- Distribution
- US distribution to states of FL, MT, NY, OH, GA, and Puerto Rico
- Location
- Lancaster, NY
Frequently Asked Questions
What product was recalled? ▼
Probe Cover, Lite poly, 6"x48", w two bands- Ultrasonic Transducer Cover Item ID: A-AU-1002N. Recalled by Exact Medical Manufacturing, Inc.. Units affected: 5621 units.
Why was this product recalled? ▼
No 510k for the product to be used in a natural or surgical opening to the body
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on June 30, 2021. Severity: Moderate. Recall number: Z-1913-2021.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11