PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported June 15, 2016

Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350 R2R, Universal SlingBar" 450, Universal SlingBar" 450 QRH, Universal SlingBar" 450 R2R, Universal SlingBar" 600 Universal SlingBar" 600 QRH Component of Liko Lifts Product Usage: Universal SlingBar 350 It is commonly used for lifting children. Universal SlingBar 350 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R

The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bo…

Recall #
Z-1916-2016
Affected scope
163,561
Initiated
March 8, 2016
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Hill-Rom, Inc. recalled Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Uni… - a moderate-severity action.

Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Uni… was recalled by Hill-Rom, Inc. in June 15, 2016. Reason: The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the…. Check the official notice for the remedy. Verify recall #Z-1916-2016 with the FDA Devices before acting.

The recall

Hill-Rom, Inc. issued this moderate-severity FDA Devices recall-The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the….

Moderate
severity level
164K units
affected scope
Class II
classification
June 15, 2016
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1916-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1916-2016) was formally reported on June 15, 2016, with the manufacturer initiating the action on March 8, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Hill-Rom, Inc. is listed as the recalling firm, operating out of Batesville, IN. Federal records list the affected scope as 163,561, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bolt is weakened there is a potential risk… Distribution data in the federal record shows the product reached: Worldwide and US Nationwide Distribution AE, AO, AR , AT , AU , AZ , BD , BE , BG , BH , BM , BN , BR , CA , CH , CL , CN , CO , CS , CZ , DE , DK , EG , ES , FI , FR , GB , GR , HK , HR , HU , ID , IE , IL , IN , IR …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

163,561

Related Recalls

6

3 from same agency

Product description

Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350 R2R, Universal SlingBar" 450, Universal SlingBar" 450 QRH, Universal SlingBar" 450 R2R, Universal SlingBar" 600 Universal SlingBar" 600 QRH Component of Liko Lifts Product Usage: Universal SlingBar 350 It is commonly used for lifting children. Universal SlingBar 350 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R). Universal SlingBar 450 Is appropriate in most cases and with most slings. Is installed on most lifts on delivery (does not apply to Viking XL). Universal SlingBar 450 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R). Universal SlingBar 600 It is intended for use with the ComfortSling Plus mod 300/310 and 350/360, or with slings the intention of which is to overlap the leg supports, Universal SlingBar 600 is available with two assembly options - fixed assembly and Quick-release Hook (QRH).

Reason for recall

The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bolt is weakened there is a potential risk for breakage with the result of a free fall of the patient. This hazard could cause potentially minor to catastrophic injuries to a patient.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1916-2016
Date reported June 15, 2016
Date initiated March 8, 2016
Recalling firm Hill-Rom, Inc.
Firm location Batesville, IN
Affected scope 163,561
Distribution Worldwide and US Nationwide Distribution AE, AO, AR , AT , AU , AZ , BD , BE , BG , BH , BM , BN , BR , CA , CH , CL , CN , CO , CS , CZ , DE , DK , EG , ES , FI , FR , GB , GR , HK , HR , HU , ID , IE , IL , IN , IR , IS , IT , JO , JP …

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

163,561 units affected - multi-state distribution scale.

Regional (<10K units) -
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1916-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350 R2R, Universal SlingBar" 450, Universal SlingBar" 450 QRH, Universal SlingBar" 450 R2R, Universal SlingBar" 600 Universal SlingBar" 600 QRH Component of Liko Lifts Product Usage: Universal SlingBar 350 It is commonly used for lifting children. Universal SlingBar 350 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R). Universal SlingBar 450 Is appropriate in most cases and with most slings. Is installed on most lifts on delivery (does not apply to Viking XL). Universal SlingBar 450 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R). Universal SlingBar 600 It is intended for use with the ComfortSling Plus mod 300/310 and 350/360, or with slings the intention of which is to overlap the leg supports, Universal SlingBar 600 is available with two assembly options - fixed assembly and Quick-release Hook (QRH).. Recalled by Hill-Rom, Inc.. Units affected: 163,561.
Why was this product recalled?
The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bolt is weakened there is a potential risk for breakage with the result of a free fall of the patient. This hazard could cause potentially minor to catastrophic injuries to a patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 15, 2016. Severity: Moderate. Recall number: Z-1916-2016.
Where was the recalled product distributed?
Distribution: Worldwide and US Nationwide Distribution AE, AO, AR , AT , AU , AZ , BD , BE , BG , BH , BM , BN , BR , CA , CH , CL , CN , CO , CS , CZ , DE , DK , EG , ES , FI , FR , GB , GR , HK , HR , HU , ID , IE , IL , IN , IR , IS , IT , JO , JP , KR , KW , LB , LI , LK , LT , LU , LV , LY , MA , MT , MX , MY , NL , NO , NZ , OM , PH , PK , PL , PT , QA , RO , RS , RU, SA , SE , SG , SI , SK , SY , TH , TN , TR , TW , UA , US , UY and ZA..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1916-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 15, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.