Severity
Moderate
FDA Devices recall · Reported August 14, 2013
Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the…
SpineFrontier, Inc. recalled Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instru… - a moderate-severity action.
Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instru… was recalled by SpineFrontier, Inc. in August 14, 2013. Reason: Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the…. Check the official notice for the remedy. Verify recall #Z-1926-2013 with the FDA Devices before acting.
The recall
SpineFrontier, Inc. issued this moderate-severity FDA Devices recall-Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1926-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1926-2013) was formally reported on August 14, 2013, with the manufacturer initiating the action on March 14, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. SpineFrontier, Inc. is listed as the recalling firm, operating out of Beverly, MA. Federal records list the affected scope as 8.
The documented reason for this recall is: Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in pl… Distribution data in the federal record shows the product reached: Distribution only in Texas.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
8
Related Recalls
6
3 from same agency
Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature that engages and holds the screw in place. There is also a feature used to lock the screw to the inserter before insertion of the screw. They can also be used to remove screws or back screws out. Screw inserters are used during the implantation of spinal pedicle screws intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Used in conjunction with Pedicle Screw Spinal system marketed as KRD1 PedFuse Pedicle Screw System .
Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in place the locking sleeve can also slide off the inserter.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1926-2013 |
| Date reported | August 14, 2013 |
| Date initiated | March 14, 2013 |
| Recalling firm | SpineFrontier, Inc. |
| Firm location | Beverly, MA |
| Affected scope | 8 |
| Distribution | Distribution only in Texas. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 14, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.