Severity
Moderate
FDA Devices recall · Reported May 30, 2018
Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to crackin…
Getinge US Sales LLC recalled ALM XTEN Surgical Light intended to provide visible illumination for the surgical field o… - a moderate-severity action.
ALM XTEN Surgical Light intended to provide visible illumination for the surgical field o… was recalled by Getinge US Sales LLC in May 30, 2018. Reason: Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing l…. Check the official notice for the remedy. Verify recall #Z-1927-2018 with the FDA Devices before acting.
The recall
Getinge US Sales LLC issued this moderate-severity FDA Devices recall-Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing l….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1927-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1927-2018) was formally reported on May 30, 2018, with the manufacturer initiating the action on November 16, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Getinge US Sales LLC is listed as the recalling firm, operating out of Wayne, NJ. Federal records list the affected scope as 667.
The documented reason for this recall is: Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams t… Distribution data in the federal record shows the product reached: U.S. and Gov. accts.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
667
Related Recalls
6
6 from same agency
ALM XTEN Surgical Light intended to provide visible illumination for the surgical field or for the examination of the patient. Part Numbers: ARD567801093, ARD567801094
Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into breaks with heavy use and aging of the equipment.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1927-2018 |
| Date reported | May 30, 2018 |
| Date initiated | November 16, 2017 |
| Recalling firm | Getinge US Sales LLC |
| Firm location | Wayne, NJ |
| Affected scope | 667 |
| Distribution | U.S. and Gov. accts. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Two federal-filing-derived peer sets for this notice, both outside the same recalling firm and filing state so the neighborhoods are not firm or geographic containment.
Same archive severity grade (Moderate), closest by report date among other firms and states.
Same product category, closest by report date among other firms and states (deduped against the severity set).
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Showing 5 of 40909 other recalls in this category.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 30, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.